In the current state of technology, the ventricular catheter connected to an external strain gauge is the most accurate, low-cost, and reliable method of monitoring intracranial pressure (ICP). It also can be recalibrated in situ. ICP transduction via fiberoptic or micro strain gauge devices placed in ventricular catheters provide similar benefits, but at a higher cost.
Parenchymal ICP monitors cannot be recalibrated during monitoring. Parenchimal ICP monitors, using micro strain pressure transducers, have negligible drift. The measurement drift is independent of the duration of monitoring.Subarachnoid, subdural, and epidural monitors (fluid coupled or pneumatic) are less accurate.
In patients for whom ICP monitoring is indicated, a decision must be made about what type of monitoring device to use. The optimal ICP monitoring device is one that is accurate, reliable, cost effective, and causes minimal patient morbidity.
The Association for the Advancement of Medical Instrumentation (AAMI) has developed the American National Standard for Intracranial Pressure Monitoring Devices in association with a Neurosurgery committee. The purpose of this standard is to provide labeling, safety, and performance requirements, and to test methods that will help assure a reasonable level of safety and effectiveness of devices intended for use in the measurement of ICP. According to the AAMI standard, an ICP device should have the following specifications:
Current ICP monitors allow pressure transduction by external strain, catheter tip strain gauge, or catheter tip fiberoptic technology. External strain gauge transducers are coupled to the patient's intracranial space via fluidfilled lines whereas catheter tip transducer technologies are placed intracranially. There is evidence that external strain gauge transducers are accurate.1 They can be recalibrated, but obstruction of the fluid couple can cause inaccuracy. In addition, the external transducer must be consistently maintained at a fixed reference point relative to the patient's head to avoid measurement error.
Micro strain gauge or fiberoptic devices are calibrated prior to intracranial insertion and cannot be recalibrated once inserted, without an associated ventricular catheter. Consequently, if the device measurement drifts and is not recalibrated, there is potential for an inaccurate measurement.
For this update, Medline was searched from 1996 through April of 2006 (see Appendix B for search strategy), and results were supplemented with literature recommended by peers or identified from reference lists. Of 39 potentially relevant studies, 7 were added to the existing tables and used as evidence for this question (see Evidence Tables I and II).
The scientific discussion of ICP monitoring technology is divided into the following sections:
A. ICP monitoring device accuracy and reliabilityA. ICP Monitoring Device Accuracy and Reliability
As specified in the Methods section of this document, the strongest evidence for the accuracy and reliability of ICP monitors would be derived from well designed studies that compare simultaneous readings from the monitor being tested to those of an established reference standard and that, among other things, would include large samples of broad-spectrum patients. The ventricular fluid coupled ICP monitor is the established reference standard for measuring ICP. Fourteen publications were identified that simultaneously compared the ventricular monitor to other monitors in a total of 273 patients with TBI (see Evidence Table I). Location of pressure transduction devices varied across studies. Sample sizes for the individual studies ranged from five to 51 patients. Due to changes in technology, only more current publications were considered relevant.
Four studies compared readings from the reference monitor to those of parenchymal strain gauge catheter tip pressure transducer device. Of those, two were published since 1995 one of which indicated that readings from the parenchymal strain gauge device varied within 2 mm Hg from those of the reference standard.
In four studies that compared readings from the reference monitor to those of parenchymal fiberoptic catheter tip pressure transduction devices only one was published since 1995, and reported a strong correlation between initial parenchymal and ventricular measurement.
Precision of parenchymal ICP monitors has also been assessed by comparing the measurement value at the time of ICP monitor removal with zero atmosphere (degree of difference = drift). Data from eight studies published since 1995 are presented in Evidence Table II. Of these, two publications report accuracy for the micro strain gauge transducer and six for the fiberoptic. However, the literature on fiberoptic transducers is outdated, as there were significant improvements for the fiberoptic transducer in the manufacturing and testing processes in 1999 (manufacturer correspondence), and studies were conducted with data collection from populations treated before the improvements were made. In 153 separate parenchymal ICP probe measurements there were less than 1% of readings above or below 5 mm Hg, when compared to zero atmosphere, at the time of the ICP device removal.
B. Optimal Intracranial Location of Monitor
A pressure transduction device for ICP monitoring can be placed in the epidural, subdural, subarachnoid, parenchymal, or ventricular location. Historically, ventricular ICP is used as the reference standard in comparing the accuracy of ICP monitors in other intracranial compartments. The potential risks of catheter misplacement, infection, hemorrhage and obstruction have led to alternative intracranial sites for ICP monitoring.
The following statements regarding ICP monitor location are derived from the primarily Class III evidence included in this review:
C. Complications
ICP monitoring complications include infection (see Infection Prophylaxis topic), hemorrhage, malfunction, obstruction, or malposition. While the current literature suggests these complications generally do not produce long term morbidity in patients, they can cause inaccurate ICP readings, and they can increase costs by requiring replacement of the monitor.
i. Hemorrhage. Hemorrhage associated with an ICP device is not defined in the majority of reports reviewed in terms of volume of hematoma on head CT, or in terms of morbidity. There were eight publications on ventriculostomy associated hematomas reporting an average incidence of 1.1% versus an article on subarachnoid bolts (no hematomas), subdural catheters (no hematomas), and micro strain gauge devices (three hematomas in 28 patients, 11%). There have been no publications on the complication rate of an improved fiberoptic transducer in populations studied since 1999. Significant hematomas receiving surgical evacuation occurred in 0.5% of patients in published reports with more than 200 patients receiving ICP monitoring.
ii. Malfunction. Malfunction or obstruction in fluid coupled ventricular catheters, subarachnoid bolts, or subdural catheters has been reported as 6.3%, 16%, and 10.5% respectively. In reports of ventricular catheter malposition, 3% of patients needed operative revision. There have been no publications on the complication rate of an improved fiberoptic transducer in populations studied since 1999. Malfunctions of micro strain gauge devices are reported as 0%.
As delineated above, each type of pressure transduction system and intracranial location of the monitor has a profile of potential complications. Calibration, monitoring for infection, and checking fluid coupled devices for obstruction are necessary tasks in maintaining an optimal ICP monitoring system. Table 2 below summarizes each type of ICP monitor by the parameters discussed above.
D. Cost
Estimated costs of the various ICP devices are presented in Tables 1 and 2. The non-disposable hardware that need to be purchased with fiberoptic and strain gauge catheter tip ICP devices range in cost from $6,000 to $10,000 per bed. ICP transduction with an external strain gauge costs $208 versus an average of $545 for micro strain gauge or fiberoptic transducers.
In patients who receive ICP monitoring, a ventricular catheter connected to an external strain gauge transducer is the most accurate and cost effective method of monitoring ICP. Clinically significant infections or hemorrhage associated with ICP devices causing patient morbidity are rare and should not deter the decision to monitor ICP.
Parenchymal transducer devices measure ICP similar to ventricular ICP pressure but have the potential for measurement differences due to the inability to recalibrate. These devices are advantageous when ventricular ICP is not obtained or if there is obstruction in the fluid couple. Subarachnoid or subdural fluid coupled devices and epidural ICP devices are currently less accurate.
| Reference | Data Class | Description of Study | Conclusion |
|---|---|---|---|
| Artru et al., 1992 | A prospective study of parenchymal fiberoptic catheter tip ICP monitors in 100 patients | Daily baseline drift of 0.3 mm Hg | |
| Barlow et al., 1985 | Simultaneous recording of ventricular fluid coupled ICP compared to a subdural fluid coupled catheter in 10 patients and a subdural catheter tip and pressure transducer device in another 10 patients. | Compared to ventricular ICP, 44% of the subdural fluid coupled device measurements 72% of the subdural catheter tip pressure transducer devices were within a 10 mm Hg range. | |
| Bavetta et al., 1997 | A prospective study of 101 fiberoptic pressure transducers (52 subdural and 42 ventricular) in 86 patients. | An average of 3.3 mm Hg zero drift was noted each day up to 5 days after insertion. 10% of devices had functional failure. | |
| Bruder et al., 1995 | Comparison of an epidural ICP monitor and a parenchymal fiberoptic catheter tip ICP monitor in 10 severe head injury patients. | There was a lack of measurement agreement with the epidural ICP on average 9 mm Hg higher (range, 10-28 mm Hg) than parenchymal ICP. | |
| Chambers et al., 1993 | Simultaneous recording of ventricular fluid coupled ICP compared to a fiberoptic catheter tip pressure transducer device at the tip of the ventricular catheter in 10 patients. | 60% of the ICP readings with the fiberoptic device were within 2 mm Hg of the ventricular fluid coupled ICP readings. | |
| Chambers et al., 1990 | ICP recordings between a ventricular fluid coupled system in 10 patients compared to a subdural fiberoptic catheter tip pressure transducer and the same device situated in the ventricular catheter in another 10 patients. | 54% and 74% of the fiberoptic subdural and fiberoptic ventricular ICP readings respectively were with 5 mm Hg of the ventricular fluid coupled ICP measurements. | |
| Czech et al., 1993 | Comparison of simultaneous ICP recordings in 15 patients using a ventricular flid coupled ICP monitoring system and an epidural pneumatic ICP monitoring device. | In the majority of comparisons the epidural device ICP measurements were different from ventricular ICP recordings with deviations between 20 and +12 mm Hg. | |
| Dearden et al., 1984 | Assessment of ICP measurement accuracy in a subarachnoid/subdural fluid coupled bolt device using an infusion test in 18 patients. | Device read ICP accurately according to infusion test 48% of the time. | |
| Gambardella et al., 1992 | Comparison of a parenchymal fiberoptic catheter tip pressure transduction device to ventricular fluid coupled ICP readings in 18 adults patients. | 55% of parenchymal fiberoptic ICP readings were 5 mm Hg higher or lower than ventricular ICP measurements. | |
| Gopinath et al., 1995 | Evaluation of the measurement accuracy and drift of a new catheter tip strain gauge ICP device. The device was placed in the lumen of a ventricular catheter in 25 patients. | No significant measurement drift v four days. The device was 63% accurate (within 2 mm Hg) compared to ventricular ICP recordings. | |
| Gray et al., 1996 | Comparison of ICP readings in 15 patients using catheter tip strain gauge devices simultaneously in parenchymal and subdural locations. | ICP measurement differences of >4 mm Hg were noted in 30% of the readings. Daily baseline drift of 0.3 mm Hg in parenchymal location. | |
| Mendelow et al., 1983 | Simultaneous recordings of ICP using two types of subdural fluid coupled bolt devices and a ventricular catheter fluid coupled system in 31 patients. | ICP recordings were within 10 mm Hg of ventricular ICP in 41% of the recordings using one type of bolt and 58% using the other kind. | |
| Mollman et al., 1988 | Simultaneous recordings of ICP using a subdural/subarachnoid fluid coupled catheter and a ventricular fluid coupled catheter in 31 patients. | The difference between the ICP readings was -0.12 mm Hg with a standard deviation of 5.29 mm Hg. | |
| Ostrup et al., 1987 | Comparison of ICP readings between a parenchymal fiberoptic catheter tip pressure transducer device and ventricular fluid coupled catheter or subarachnoid bolt in 15 adults and children. | Measurement drift up to 1 mm Hg per day. Parenchymal ICP readings were generally within 2-5 mm Hg of ventricular or 5 subarachnoid ICP measurements. | |
| Piek et al., 1990 | In a series of 100 patiens, 13 had simultaneous ICP recordings from a parenchymal strain gauge catheter tip pressure transducer device and a ventricular fluid coupled catheter. | An initial drift up to 4 mm Hg in the first day. Parenchymal ICP measurements were generally 4-8 mm Hg below ventricular ICP. | |
| Piek et al., 1987 | Simultaneous recordings of ICP using a parenchymal strain gauge catheter tip pressure transducer device and a ventricular fluid coupled catheter in seven patients. | Parenchymal ICP was 4-12 mm Hg lower than ventricular ICP but parallel changes in pressure were noted. | |
| Powell et al., 1985 | Simultaneous recordings of ICP using an epidural pneumatic pressure transducer and a ventricular fluid coupled catheter in 17 patients. | Marked differences in pressure up to 30 mm Hg were recorded. | |
| Schickner et al., 1992 | Comparison of ICP readings between a parenchymal fiberoptic catheter tip pressure transducer device and ventricular fluid coupled catheter in 10 patients. | 66% of the parenchymal fiberoptic measurements exceeded ventricular ICP and 21% were lower. Absolute pressure differences of up to 40 mm Hg were recorded. | |
| Schwartz et al., 1992 | Comparison of ICP readings between an epidural pneumatic pressure transducer device and a subdural strain gauge, subdural fiberoptic or ventricular fluid coupled catheter 6 patients. | ICP readings from the epidural device correlated with the other device readings in only one case. | |
| Shapiro et al., 1996 | Review of clinical performance of parenchymal fiberoptic catheter tip ICP monitors in 244 patients (180 head injury) of which 51 also had ventricular catheter placement. | A strong correlation was found between initial parenchymal and ventricular measurements. Fiberoptic breakage and malfunction was seen in 17% and 14% of patients, respectively. The mean length of monitoring was 7 days. | |
| Weaver et al., 1982 | Comparison of ICP measurements between two subarachnoid fluid coupled pressure transducers in the same patient. Twenty patients were studied, four of them had unilateral mass lesions. | More than 50% of patients demonstrated significant differences in ICP. Patients harboring intracranial mass lesions showing clear differences. | |
| Koskinen et al., 2005 | A prospective study in 28 patients with parenchymal micro strain gauge ICP transducer and in 22 patients with parenchymal microstrain gauge ICP transducers and concurrent ventriculostomies. | Only 21% of the probes showed zero drift greater than ±2 mm Hg when removed. 22% of the probes read more than ±2 mm Hg compared to ventricular CSF pressure readings. Three hematomas (nonoperable) and no significant infections (probes were not cultured). | |
| Martinez-Manas et al., 2000 | Prospective study done in 1997 of 101 patients (71% TBI) all patients had GCS < 9 who had 108 consecutive fiberoptic ICP monitors placed (63% parenchymal, 28% subdural and the rest intraventricular. | Probe tips were sent for culture and 13.2% were positive. Intracranial hematoma occurred near the probe placement in 4%. 89% of the probes showed a positive or negative drift after removal (range -24 to +35 mm Hg which was not correlated with duration of monitoring. | |
| Munch et al., 1998 | Parenchymal (n = 104) and ventricular (n = 32) fiberoptic transduced ICP devices were placed. Accuracy of expected ICP was assessed by neurological exam and CT scan. 118 patients studied prospectively over an 18-month period. Fiberobtics (104) and ventrics (32) placed. Reliability assessed by neuro exam and CT, complications assessed. | 85% of the ICP devices were deemed reliable. Complications included 18.1% needed replacement due to failure. 23.5% were dislocated. Only one positive CSF culture noted. | |
| Piper et al., 2001 | Zero drift characteristics of 34 parenchymal fiberoptic probes studied in 50 patients with a 4-day mean duration of ICP monitoring (range 1-12 days) | 50% of the parenchymal probes had measurements greater than ±3 mm Hg after removal when compared to zero drift. There was no correlation with the duration of monitoring. | |
| Poca et al., 2002 | 163 patients who had 187 fiberoptic parenchymal bolts placed prospectively and studied over a three year period. All patients had TBI and GCS < 9. Mean duration of monitoring was 5 ± 2.2 days. | 89% of probes showed drift (-12 to +7 mm Hg) when removed and 17% had positive culture of the probe tip. 10% sensor malfunction and 2.8% hematoma rate (nonoperable) was reported. | |
| Signorini et al., 1998 | 10 patients (8 TBI) had placement of micro strain gauge parenchymal ICP monitor and comparisons with fiberoptic parenchymal monitors (5) and intraventricular fluid coupled monitors (5) were performed. | A difference of 9 mm Hg was noted between the two parenchymal monitors. Following removal, 33% of the micro strain gauge monitor readings and 50% of the fiberoptic monitor readings were greater than ±2 mm Hg from zero drift, respectively. | |
| Stendel R et al., 2003 | Prospective comparison testing of the Neurovent ICP and fiberoptic parenchymal probes in 148 patients (72% TBI) of whom an early group of 50 patients received fiberoptic probes and then 98 had Neurovent parenchymal monitors placed. | Hematomas were noted in 2% and 1% of fiberoptic (C) and Neurovent (N) probes respectively. Technical problems in the following: dislocation 14% (C) and 2% (N), damage 6% (C) and 5% (N), Error 8% (C) and 0% (N) and drift 3.5 mm + 3.1 (C) and 1.7 mm + 1.36 (N) were reported. |